The pH scale typically ranges from 0 to 14. According to the Nernst equation pH values are affected by temperature, hence the temperature compensation is a vital function of any pH Meter to ensure accurate pH measurement.
Technical Specifications
Detailed operating and construction parameters for this model.
Model |
BMET-201 |
| pH | |
| – Range | -2.00 to 14.00 pH |
| – Resolution | 0.01 pH |
| – Accuracy | ± 0.01 pH |
| – Calibration Points | Up to 2 |
| – Standard Recognition | NIST buffers |
| mV | |
| – Range | -1800 to 1800 mV |
| – Resolution | 1 mV |
| – Accuracy | ± 0.1 %FS |
| Conductivity | |
| – Range | / |
| – Resolution | / |
| – Accuracy | / |
| – Reference Temperature | / |
| Measurement | |
| – Reading Mode | Continuous |
| – Reading Prompts | Reading, Stable |
| – Temp. Compensation | MTC |
| Inputs | |
| – pH Electrode | BNC(Q9) |
| – Conductivity Probe | / |
| Display Options | |
| – Backlight | Yes |
| – Auto Shutdown | 300, 600, 1200, 1800, 3600 sec., off |
| – IP Rating | IP54 |
| General | |
| – Power | AC Adapter, 100-240 V AC input, DC 9V output |
| – Dimensions | 200x160x63 mm |
| – Weight | 600 g (1.32 lb) |